Biosafety cabinets after a move: why recertification is needed

Why a biosafety cabinet (BSC) needs recertification after it is moved, what is tested, and how the lab can plan the test to avoid delays.

Illustration: Biosafety cabinets after a move: why recertification is needed 1.8 m

In short

Moving a biosafety cabinet can disturb HEPA filter seals and airflow, so common standards require recertification before use. Testing is the lab's responsibility: book a certifier for the days after placement and keep the cabinet out of use until then.

Why it is required

A BSC protects the worker, the sample and the environment with precise airflow and HEPA filters. Vibration during transport can shift a filter or break a seal without any visible sign, so it is tested after every move.

What is usually tested

  • HEPA filter integrity.
  • Airflow velocity and direction.
  • Alarms and fan operation.
  • Exhaust connection, if any.

Avoid delays

  • Book the certifier when you set the moving date.
  • Make sure the cabinet is placed and connected before the certifier arrives.
  • No certifier yet? We are happy to give you the phone number of a qualified one we know.
  • If you have several cabinets, move them in phases.