Cleanroom ISO classes: how to read ISO 14644

ISO 14644 cleanroom classes explained: particle limits for each class, the old Class 100 names, EU GMP grades A–D, and when a cleanroom has to be tested again.

Illustration: Cleanroom ISO classes ISO 14644-1 ISO 8ISO 7ISO 6ISO 5 3,520

In short

A cleanroom's class is set under ISO 14644-1, based on how many particles are allowed in a cubic metre of air. The lower the number, the cleaner the room: ISO 5 matches the old US Class 100, and ISO 7 matches Class 10,000. Testing is done by a qualified tester working for the room's owner, and the room is tested again after significant changes.

What ISO 14644 is

ISO 14644 is the international series of standards for cleanrooms and controlled environments. Part 1, ISO 14644-1, sets how air cleanliness is classified by particle concentration. Part 2, ISO 14644-2, covers monitoring: how you show over time that the room still meets its class. Both parts were revised in 2015.

The classification table: how many particles each class allows

Classes run from ISO 1, the cleanest, to ISO 9. Each class allows ten times as many particles as the one before it. The table shows the maximum number of particles of 0.5 µm and larger per cubic metre of air, the size most often used for comparison, and the matching name in the old US standard.

ISO classMax. particles per m³ (0.5 µm and larger)Old US name
ISO 335Class 1
ISO 4352Class 10
ISO 53,520Class 100
ISO 635,200Class 1,000
ISO 7352,000Class 10,000
ISO 83,520,000Class 100,000
ISO 935,200,000—

The standard also has columns for other sizes, from 0.1 to 5 µm, and each room is classified at the sizes specified for it. A full classification therefore states the particle size and the occupancy state at the time of testing, for example: ISO 7, at rest, at 0.5 µm.

Class 100 and Class 10,000: where the old names come from

Before ISO 14644, cleanrooms were classified under the US standard FED-STD-209E. It counted particles of 0.5 µm and larger per cubic foot, and the class name was the maximum count: Class 100 allows up to 100 particles per cubic foot, about 3,520 per cubic metre, which is ISO 5. The standard was cancelled in November 2001 and replaced by ISO 14644, but many people and suppliers still say Class 100 or Class 10,000, so it helps to know both systems.

EU GMP grades A, B, C and D

Pharmaceutical plants and sterile manufacturing work to EU GMP Annex 1, which uses grades A to D. Each grade is defined in two states: at rest and in operation. By the concentration of 0.5 µm particles, the grades line up with these ISO classes:

GradeAt restIn operationTypical use
AISO 5ISO 5Critical zone, e.g. aseptic filling
BISO 5ISO 7Background for grade A
CISO 7ISO 8Preparing solutions before sterilisation
DISO 8Set by risk assessmentWashing and preparing equipment and components

Which state is tested: as-built, at-rest, operational

The standard distinguishes three occupancy states:

  • As-built: the room is complete and the air systems are running, but there is no equipment and no one inside.
  • At-rest: equipment is installed, but there is no one in the room.
  • Operational: the room is working normally, with its equipment and the agreed number of people.

The same room can meet ISO 5 with no one inside and fail it during work, because people and movement add particles. That is why a classification has to state which state the room was tested in. New equipment brought into the room also changes the picture, which is one reason rooms are tested again after a move.

When the room is tested again

Under ISO 14644-1, classification is repeated at intervals set by risk assessment, typically once a year. In rooms that work to Annex 1, grades A and B are requalified at least every 6 months, and grades C and D at least every 12 months.

Besides periodic testing, Annex 1 also calls for requalification after changes, for example:

  • An interruption of airflow that affects how the room operates.
  • A change to the room's design or to the HVAC settings.
  • Special maintenance, such as replacing the final filters.

Testing usually covers particle counts, filter integrity, airflow volume and velocity, and pressure differences between rooms. A qualified tester working for the room's owner carries it out, and QA decides which tests are needed after each change.

What this means when equipment moves in or out

The class also sets how the move is done. The cleaner the room, the stricter the procedure: how equipment is cleaned and wrapped before it goes in, which airlock it passes through, what the crew wears and how many people may enter. Before a move we ask the cleanroom team for the class and the entry procedure, and plan packing, crew and sequence around them. The steps themselves are in our guide Moving equipment into and out of a cleanroom.

Every door opening and every item brought in raises the particle count. Leave time in the schedule for cleaning and testing before work resumes, and if the air system is switched off or the room is altered during the works, book requalification in advance as well.

What the customer is responsible for

We plan the move, pack to the room's procedure, and transport and position the equipment. Cleaning the room, testing and classification are done by your QA team or by a qualified testing company you appoint, and we coordinate the schedule with them so testing can start right after positioning.